| Probe Specifications: |
|
FOXY Formulation |
FOSPOR Formulation |
HIOXY Formulation |
| |
(general-purpose coating) |
(high-sensitivity for low levels of O2) |
(for hydrocarbon environments) |
| O2% range (at 1 ATM): |
0-100% |
0-10% |
0-20.9% |
| DO range (ppm at 1 ATM): |
0-40 ppm |
0-4 ppm |
0-8 ppm |
| Temperature range: |
-50-+80 oC for probes |
-50-+80 oC for probes |
-50-+60 oC |
| 0-60 oC for patches |
0-60 oC for patches |
NA |
| O2% resolution: |
0.01-0.05% |
.001% |
0.01-0.05% |
| DO resolution (at room temp): |
4-20 ppb |
0.4-2 ppb |
4-20 ppb |
| O2% accuracy: |
5% of reading |
5% of reading |
5% of reading |
| DO accuracy: |
5% of reading |
5% of reading |
5% of reading |
| Min. detectable level in gas: |
0.01% - 0.05% |
0.001% - 0.01% |
0.01% - 0.05% |
| Response time: |
< 1 second in gas |
< 1 second in gas |
< 1 second in gas |
| ~30 seconds with overcoating in gas |
~30 seconds with overcoating in gas |
~45 seconds in pure water |
|
~45 seconds in pure water |
~45 seconds in pure water |
|
| |
| Patch Specifications: |
| |
FOXY Formulation |
FOSPOR Formulation |
HIOXY Formulation |
| 02% range (at 1 ATM): |
0-100% |
0-20.9% |
0-100% |
| DO range (ppm at 1 ATM): |
0-40 ppm |
0-8 ppm |
0-40 ppm |
| Temperature range: |
-20 to +60 oC for patches |
-20 to +60 oC for patches |
-20 to +60 oC for patches |
| 02% resolution (20 sex averaging): |
0.05% |
0.01% |
0.05% |
| DO resolution (at room temp): |
20 ppb |
4 ppb |
20 ppb |
| 02% accuracy: |
5% of reading |
5% of reading |
5% of reading |
| DO accuracy: |
5% of reading |
5% of reading |
5% of reading |
| Min. detectable level: |
0.1% O2 |
0.01% O2 |
0.1% O2 |
| Min detectable level in water (at room temp): |
40 ppb |
4 ppb |
40 ppb |
| Response time: |
<1 second in gas |
<1 second in gas |
<1 second in gas |
| |
~30 seconds with overcoating in gas |
~30 seconds with overcoating in gas |
|
| |
~45 seconds in pure water |
~45 seconds in pure water |
~45 seconds in pure water |
| |
| Sterilization Guide: |
| |
FOXY Formulation |
FOSPOR Formulation |
HIOXY Formulation |
| Form factor: |
Probes/Patches |
Probes/Patches |
Probes |
| Autoclavable: |
No |
No |
Yes |
| Calibration: |
|
|
Single point reset required after sterilization |